Protek A333 3.2GHZ Network Analyzer
Measurement range
- Frequency Range : 0.3 to 3200MHz
- Impedance : 50Ω (75Ω)
- Test Port Connectors : N-type Female
- Number of Test Ports : 2
- Frequency Accuracy : ±5 ppm
- Frequency Resolution : 1mHz
- Number of Test Points User-Selectable : 2 to 10001
- IF bandwidth settings1/1.5/2/3/5/7/10 step : 1Hz to 30kHz
- Dynamic RangeIF Bandwidth 10Hz : 2 to 3200MHz > 125dBTyp. 130dB
Test port output
- Power range : -45 to +10dBm
- Power accuracy : <1.0dB
- Power resolution : 0.05dB
- Harmonic distortion : -30dBc
Test port input
- MatchWithout system error correction : > 25dB
- Damage level : +26dBm
- Damage DC voltage : 35V
External Reference Signal Input/Output
- Frequency : 10MHz
Measurement Accuracy
1. Accuracy of Transmission Measurement (magnitude/phase)
- Specifications are based on a matched DUT, a measurement bandwidth of
1 Hz,
and a nominal source power of -5dBm +15 to +5dB 0.2dB/2.0°
+5 to -50dB 0.1dB/1.0° , -50 to -70dB 0.2dB/2.0° , -70 to -90dB
1.0dB/6.0°
2. Accuracy of Reflection Measurement (magnitude/phase)
- Specifications are based on a matched DUT, a measurement bandwidth of
1 Hz, and a nominal source power of -5dBm
0 to -15dB 0.4dB/4.0° , -15 to -25dB 1.5dB/7.0° , -25 to -35dB 4dB/22.0°
Trace Stability
Temperature DependencePer one degree of the temperature variation :
0.02dB
Trace Noise MagnitudeIF bandwidth 3kHz : 0.001dB rms
Other capabilities
1. Internal hard disk drive
Store and recall instrument states, calibration data, and trace data on
internal hard drive. Trace data can be saved in CSV (comma separated
value) format. Instrument states include all control settings, limit
lines, segment sweep tables, and memory trace data.
2. USB
Instrument states, calibration data, and trace data can be stored on an
external USB drive.
3. Screen hardcopy
Printouts of instrument data are directly produced on a printer. The
analyzer provides USB and parallel interfaces.
4. Familiar graphic user interface
Protek A333 provides graphic user interface based on Windows. There are
three ways to operate A333 manually: a hard-key interface, a touch
screen interface, mouse interface.
5. Limit lines
Defining the test limit lines that appear on the display for pass/fail
testing. Defined limits may be of any combination of horizontal/sloping
lines and discrete data points.
Programming functions
COM/DCOM Automation
Remote Control
LAN, GPIB (optiona)
Operating environment
Temperature : +5 to +40°c
Dimension ( W x H x D ) : 426 x 222 x 270 mm
Weight : 10.6 kg
Power supply : 100 to 240 VAC / 47 to 63 Hz
A333 network analyzer in 3.2GHz (all software option included but frequency offset)
| Protek A333 Data Sheet | Price: USD $22,000.00 |
| Availability 4 weeks ARO (after receipt of order) | |